312 Portable GC

 

Specifications


Detectors:


Photoionization detector (PID), Flame ionization detector (FID), Thermal conductivity detector (TCD), Flame photometric detector (FPD), Far UV absorbance (FUV) detector, Catalytic combustion detector (CCD)


Species measured:


PID & FID: Organics (VOCs by PID or FID) and inorganic species (H2S, NH3, AsH3… ) that can be ionized by the PID lamp (9.5, 10.6 or 11.7 eV)
TCD: Universal detector- fixed gases (O2, N2, CO…), organic and inorganic compounds, very sensitive for H2, He, Xe
FPD: Selective response for S or P compounds
Far UV: Nearly universal response for organic and inorganic compounds
CCD: Combustible compounds such as hydrocarbons


Detection Limit:

PID- 0.5 ppb benzene
FID- 50 ppb benzene
TCD- H2- 10 ppm; O2 150 ppm


Linear Dynamic Range:


PID- 108
FID- 2 x 106
TCD- 2 x 104
FPD- S- 103; P 105
Far UV- 106
CCD- 5 x 104


Sensor(s):


1 or 2 detectors


Precision:


+/- 1-2% of reading


Response time:


Depends on retention times
Sample flow rate:
> 0.5 LPM


Injection:


On-column syringe injection or automatic with 6 port gas valve (optional)
Span Drift:
<1% over 24 hours


Zero Drift:


NA- auto zero on each run
Power consumption:
< 50 watts maximum; 115 VAC, 0.5 amps (12 VDC 5 amps???)
Dimensions:
enclosure - 18.75''x 13.75'' x +7.5''


Weight:


Fiberglass case - 27.5 pounds
Instrument operating conditions:
5-40 degrees C, 0-95% RH (non-condensing)


________________________________________


Features


PID Lamps


Our 9.5 and 10.6 eV lamps remain the best in the business since we developed them over 30 years ago.


FID


Our very sensitive FID has flame out sensor that will shut off the hydrogen if the flame goes out


FPD


Our very sensitive FPD has flame out sensor that will shut off the hydrogen if the flame goes out


Display


Our 7'' Color VGA display (640 x 480 resolution) is easy to read. The display is a touch screen


PC


Embedded Pentium Computer with > 1 GByte of RAM; data storage on 1 G byte USB memory stick
Operating system
Windows XP® embedded stored on 2.5 GByte smart card (the Peak Works program and all data are on the memory stick


Peak Detection & System Operation:


Controlled by 32 bit Peak Works software program with more than 20 years of development
ADC
Our 16 bit ADCs provide a resolution of 1 part in 64,000 and our signal algorithm minimizes noise by signal averaging in the msec range; the ADC can acquire positive or negative signals


Auto ranging


The 312 has 3 ranges with computer controlled auto ranging- x1, x 10, x 100


Auto zero


Beginning of each run


Programmable Voltage/current output


0-1VDC


Digital outputs


USB connection


Battery life


3-4 hours; can operate off cigarette lighter adapter


Carrier gas


50 liters at 1500 PSI; typical capillary column 15-20 cc/min; typical packed column 30-40 cc/min
Oven


Size:


Typical column diameter: 3.5''


Temperature range:


Battery: 25-50 Degrees Celsius


External power:

25-100 Degrees Celsius


Programmable setpoints


The low and high set points (depends on the options) can be customer programmed for each point and/or compound

 

Contact Us to discuss service contracts, overhauls, on-site support, parts and consumables.

portable, field testing, hand held, GC, FID, PID, HNU, photoionization, infrared, electrochemistry, catalytic combustion, flame ionization, capacitance, thermal conductivity, process gas chromatography VOCs, O2, H2O, fixed gases, SO2, NO2, NO, CO2