The Thermo Scientific* iPRO 5000 NS Total Nitrogen and Total Sulfur Analyzer offers breakthrough versatility with exceptional accuracy and precision.
Specifications
|
||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Instrument Properties | ||||||||||
| Dimensions (D x W x H) | 0.53 x 0.54 x 0.37m | |||||||||
| Weight | 35kg | |||||||||
| Septum Type | Merlin Microseal* | |||||||||
| Typical Septum Lifetime | 25,000 injections | |||||||||
| Operating Temperature | 15 - 35°C | |||||||||
| Operating Humidity | 5% @ 15°C - 60% @ 35°C | |||||||||
| Furnace Operating Temperature | 900 - 1150°C | |||||||||
| Injection Module | jetPRO* Direct Injector | |||||||||
| Software | NSX Visual | |||||||||
| Pre-loaded Methods | ASTM & customizable methods | |||||||||
| Performance | ||||||||||
| Injected volume | 0 - 250µL | |||||||||
| Sample Injection Speed | 0 - 2.5µL/sec | |||||||||
| Typical Sample Analysis time | 3 min. | |||||||||
| Typical Linearity (R2) | 0.995 | |||||||||
| Working Range | 30ppb - 10,000ppm | |||||||||
| Typical RSDs | Greater of 40 ppb or 2% | |||||||||
| Method Compliance | ASTM D4629, ASTM D7184, ASTM D6069, ASTM D5453, ASTM D7183, ASTM D6667, EN ISO 20846, EN ISO 15486, UOP 936 | |||||||||
| Electrical | ||||||||||
| Voltage | 100 - 240V | |||||||||
| Frequency | 50/60 Hz | |||||||||
| Max. Power Consumption | 1500 VA | |||||||||
| Gases | ||||||||||
| Primary Oxygen | ||||||||||
| Minimum Oxygen Grade | 2.6 (99.6%) | |||||||||
| Oxygen Consumption | 600 - 800mL/min. | |||||||||
| Oxygen Input Pressure | 3 - 7barg | |||||||||
| Carrier Gas: Choice of Oxygen, Argon or Helium | ||||||||||
| Minimum Oxygen Grade | 2.6 (99.6%) | |||||||||
| Oxygen Consumption | 600 - 800mL/min. | |||||||||
| Oxygen Input Pressure | 3 - 7barg | |||||||||
| Minimum Argon Grade | 4.8 (99.998%) | |||||||||
| Argon Consumption | 0 - 200mL/min | |||||||||
| Argon Input Pressure | 3 - 7barg | |||||||||
| Minimum Helium Grade | 4.8 (99.998%) | |||||||||
| Helium Consumption | 0 - 200mL/min | |||||||||
| Helium Input Pressure | 3 - 7barg | |||||||||
| A/I 3000 Autosampler | ||||||||||
| Dimensions (D x W x H) | 0.25 x 0.44 x 0.23m | |||||||||
| Weight | 9kg | |||||||||
| Vial Size | 2mL | |||||||||
| Number of Positions | 8 | |||||||||
| A/S 3000 Autosampler | ||||||||||
| Dimensions (D x W x H) | 0.25 x 0.44 x 0.40m | |||||||||
| Weight | 10kg | |||||||||
| Number of Positions | 105 | |||||||||
| Vial Size | 2mL | |||||||||
| EGM-III Gas/LPG Introduction Module | ||||||||||
| Dimensions (D x W x H) | 0.45 x 0.25 x 0.7m | |||||||||
| Weight | 22kg | |||||||||
| LPG Loop Volume | 100µL | |||||||||
| Gas Loop Volume | 10mL | |||||||||
| Typical Analysis Time | 5 min. | |||||||||
| Number of Loops | 1/1/1999 | |||||||||
| Max Sample Pressure | 20bar | |||||||||