Specifications
Operating temperature : 4'C above ambient to 450'C*
Temperature set-point resolution : 1'C
Maximum set-point temperature rate : 100'C/min
Maximum run time : 9,999 min
Programming ramp/plateaus : 15/16
Column bleed compensation standard for two channels
Temperature Stability : +-0.05'C
Fast cooling down rates : 6.5 min(from 450'C to 50'C)
* With LN2 cryo : -80'C to 450'C
* With CO2 cryo : -55'C to 450'C
You
choose the injection method
Manual
or Automatic
Split
/ Split less
Packed column injection
Gas Sampling Valve (GSV)
Pyrolysis, Pyroprobe
Headspace, Liquid and SPME Auto sampler
Purge Housing
Sample
Preperation
Headspace,
Liquid and SPME Auto sampler
This is a single unit combining static Headspace analysis, Liquid sample
injection and SPME (Solid Phase Microextraction)! It can save your time
and money, increase analysis flexibility.
Pyrolysis,
Pyroprobe
This pyrolysis allows you to analyze the samples which were previously
unsuitable for analysis without extractions or derivatizations.
By using pyrolysis, samples such as paint, adhesives, tapes, caulk, food
packaging, rubber, plastic,
papers, ink, coating and a full range of household products can be analyzed
for qualitative and
quantitative information
Purge
Housing
When you analyze samples of O2, N2, etc with low concentration of a few
ppms, the purge housing is
very useful for this analysis. Purge housings eliminate any possible diffusion
from the atmosphere into
the valve, or safely vent fugitive emissions from the valve.
Flame Ionization Detector (FID)
Thermal Conductivity Detector (TCD)
Pulsed Discharge Detector (PDD)
Nitrogen Phosphorous Detector (NPD)
Flame Photometric Detector (FPD)
Electron Capture Detector (ECD)
Detectors
Flame
Ionization Detector (FID)
For detection of any organic compounds which can be ionized by hydrogen/air
flame
Maximum temperature : 450’C
Automatic flame ignition/flame out detection
MDL : < 3.2 pg carbon/sec as dodecane using N2 carrier and 0.4572 mm
jet
Linear Dynamic Range : < +-10 %, 107 with N2
carrier and 0.4572 mm jet
Sensitivity : 19 m Coulomb/sec
Thermal
Conductivity Detector (TCD)
For universal detection of every compound including inorganic compounds
except the carrier gas used
Maximum temperature : 400'C
Flow through cell : 4 Rhenium - tungsten filaments
MDL : < 5 ng dodecane/ml
Linear Dynamic Range : 105 (+- 5 %)
Pulsed
Discharge Detector (PDD)
PDD has two different modes, one is PDECD and the other is PDHID. The
PDECD is a selective detector for monitoring high electron affinity compounds
such as freons, chlorinated pesticides, and other halogen compounds. For
this type of compound, the minimum detectable quantity (MDQ) is at
the femtogram (10-15) or picogram (10-12)
level. The PDD is similar in sensitivity and response characteristics
to a conventional radioactive ECD, and can be operated at temperatures
up to 400'C. For operation in this mode, He and CH4 are introduced just
upstream from the column exit.
The other mode, the PDHID is an universal, non-destructive, high sensitivity
detector. The response to both inorganic and organic compounds is linear
over a wide range. Response to fixed gases is positive (increase in standing
current), with an MDQ in the low ppb range.
Nitrogen
Phosphorous Detector (NPD)
For selective detection of organic nitrogen and phosphorous compounds
Maximum temperature : 400'C
MDL : < 0.4pg N/sec(Azobenzene) or < 0.2pg P/sec(Malathion)
Dynamic Range : > 104 (N and P)
Selectivity : 35,000 to 1 : gN/gC or 75,000 to 1 : gP/gC
Flame
Photometric Detector (FPD)
For selective detection of sulfur and phosphorous compounds
Sensitivity : 5.0 x 10-13 gP/sec (P mode), 5.0
x 10-11 gS/sec (S mode)
Dynamic Range : 105 (P mode), Calibration curve is compulsory(S mode)
MDL : Vapona -1/ul Sol 15ppb Vapona in Hexane produces a peak 2x the noise
Tetrahydrothiophene -1/ul Sol 0.5ppm in Benzene gives a peak 2x the noise
Optical Filters : 525 nm(P mode), 394 nm(S mode)
Features
Powerful
Control
- Perfect control of all parameters by Autochro-3000 Data System
- Stable and precise control with powerful dual CPUs
- Remote Time Control : Remote Time Delay and Remote Signal Output
Easy
Operation
- Intuitive user interface for a keyboard
- Column conditioning function
- Stores up to 20 different analytical methods
Economical Function
- Time scheduling function
: Scheduling of system ON and OFF
- Run time based control
: Automatically repeatable runs up to 9,999 analyses
Safety Function
- Automatic flow & temperature control
: Detector “ON” after checking flow between reference and
sample to protect TCD cell
- Automatic flame ignition after checking the FID temperature
: Air/H2 flow will be blocked automatically if auto-ignition is set as
“OFF”
Data
System
The
sophisticated Clarity data system is easy to use and offers extensive
data management plus full control of the GC. The software is designed
for 21 CFR Part 11 Compliance and through full compatibility with MS Windows
XP and Vista seamlessly handles data processing and instrument control
using an ultra-reliable LAN interface.
Overlay
YL-Clarity simultaneously displays a virtually unlimited number of chromatograms
and their mathematical modification; for example,mutual deductions or
derivations of any order.
Measuring
Simultaneous data acquisition from up to four independent chromatographs,each
chromatograph can acquire data from up to 12 detectors.
|