Mercury Porosimeter Pascal
240
Pascal 240 for high pressure analysis
(up to 200 MPa)
The Pascal 240 porosimeter operates in the
classical range from atmospheric pressure up to 200 MPa. It takes
over where the Pascal 140 leaves off to measure pores down to the
lower mesopore region. This unit is versatile, high-pressure porosimeter
working up to 200 MPa. Its large autoclave can accept dilatometers
of different sizes, enabling the user to characterize samples of different
porous structures (i.e. homogeneous, heterogeneous, low or high porosity).
The accuracy and reproducibility are the utmost thanks to the PASCAL
system. It requires a sample preparation unit to degas the sample
before the analysis and to fill the dilatometer with fresh mercury.
The dilatometer preparation unit can be the Pascal 140 or a Duplex
Dilatometer filling device.
Features
- Motor driven autoclave raising/lowering
simplifies the operation and contributes to the user friendliness
of the unit
- Fast autoclave oil filling by special peristaltic
pump
- Real blank correction for a real pore volume
measurement avoiding any mercury compression and temperature effects
- Large autoclave to accept any types of dilatometer
to cover most of the applications
- Very high speed permits to collect a penetration
curve up to 200 MPa in about 9 minutes
- 4 security systems assure total safety for
the operator
- Penetrated volume detection by capacitive
electrode provides highest resolution
- Simple, easy to handle and inexpensive dilatometers
are suitable for any kind of samples
- Built-in memory data buffer stores the analysis
data freeing the PC during the run
- Multi RS232 connection permits up to 4 different
instruments to be connected to a single serial port
- Automatic calibration routine for periodical
check of the instrument status
- Separately selectable speeds to increase
and decrease pressure minimize hysteresis effects
- Extremely high resolution (up to 2500 data
point per analysis)
- Choice of operation via control panel or
PC assures high versatility of use
- 'PASCAL' pressurization method eliminates
the dead times automatically assuring equilibrium condition
- 'PASCAL' pressurization method does not
require any prior knowledge about the sample characteristics
- Stand-alone module provides extreme system
flexibility and money saving
Speed and resolution to match any requirement
The Pascal 240 incorporates a totally new pressurization
system, developed to meet the sophisticated analytical requirements
of laboratories working with modern materials. The pressurization
system uses a reversible pump operating continuously and a new type
of pressure multiplier. This permits a perfect control of the pressurization
speed during the analysis. The key features of this system are the
extremely high maximum speed of pressurization and the immediate acceleration
or deceleration response according to the real mercury penetration
speed in the sample porous matrix. At the maximum speed, the instrument
can provide a penetration curve up to 200 MPa in about 9 minutes,
permitting an incredible productivity in terms of analyses per hour,
as required in quality control labs. The easy-to-use control panel
with a liquid crystal display allows operating the unit even without
a PC connected. It lets the operator program the analysis or the calibration
and provides a useful guide to operate correctly. It shows the analysis
status in real time and indicates incorrect operations by means of
error messages. The software for instrument control and data processing
is developed in Windows 95/NT/98 © operative system.
In the Pascal series safety
is the keyword
One electromechanical, one electronic and two
mechanical protection devices assure that nothing is left to chance
and the operator’s safety is absolute. A calibrated valve in the pressure
multiplier opens if the safety pressure is reached, discharging the
oil to reduce the pressure. At the same time, a capacitive detector
continuously monitors the position of the pressure multiplier piston,
and, if the piston moves over a fixed limit, the system shuts down.
The signal of the pressure transducer connected to the autoclave is
monitored by the safety system to avoid any over-pressurization in
the dilatometer area. Finally, in case of power failure, the natural
leakage of the reversible pump assures that the pressure will decrease
by itself avoiding any risk for the operator.