Pycnomatic Technical Specifications


Standard sample chamber volumes About 20, 40 and 60 cc. (indicative maximum sample vessels volumes)


Optional sample chamber volumes Extra Small (about 4 cc) and Extra Large (about 100 cc)


Reference chamber volumes About 20, 40 and 60 cc (indicative calibrated reference chambers)


Temperature control range (only ATC) From 18 to 35 °C by a Peltier device, selectable by steps of 0.01 °C
Temperature resolution Internal reading 0.001 °C, displayed resolution 0.01 °C


Temperature sensor types Three PT100 (model ATC)
Typical temperature control stability (ATC) +/- 0.01 °C in the above temperature control range (in typical operation conditions)


Pressure transducer range From vacuum up to 250 kPa absolute reading
Pressure transducer type Absolute, piezo-resistive sensor, temperature compensated, linearized

Pressure displayed resolution 0.001 kPa


Typical pressure transducer stability +/- 0.002 kPa (in typical operation conditions)


Purging procedures By gas pulses, continuous flow or vacuum (vacuum pump not enclosed)


Calibration procedure Integrated, storing up to 3 reference chambers volume and 3 sample chamber volumes


Calibration method By certified stainless-steel spheres (spheres not enclosed)


Maximum cycles number per run 100 (user’s selectable)


Memory capacity Up to two complete runs made of up to 100 cycles each


Typical reproducibility (ATC) Better than 0.01 % on sample volume at 20 °C (evaluation on dry and thermally
equilibrated samples, sample real volume filling at about 66 % of nominal vessel
volume)


Typical accuracy (ATC) Better than 0.01 % on sample volume at 20 °C (evaluation on dry and thermally
equilibrated samples, sample real volume filling at about 66 % of nominal vessel
volume)


User’s interface Backlit display 40 characters x 4 lines, alpha-numeric keyboard
External connections Gas-in port (research grade helium or nitrogen), direct gas-out port and gas-out
through restriction.


Communication ports Serial port (9 pins) to PC and parallel to printer for reporting, serial port (9 pins) to
balance for sample weight transfer


Environment conditions Temperature: 15 to 30°C; Humidity: 20 to 80% RH; Maximum altitude: 2000 m
Power supply 100-240 VAC, 50/60 Hz, 240 VA


Dimensions Width: 25 cm , Height 33 cm, Depth: 45 cm